HumaStar 600
| User manual
114
5. If the new values are around 650 press
Save new values
to save them.
8.4.3 CLEAN SYSTEM BOTTLES
The system wash- and waste bottles need to be cleaned manually once per
month. This is not controlled by analyzer software. Build of dirt and potential
biological contamination within the containers could seriously influence perfor-
mance of the analyzer. To clean the bottles follow the below instructions.
1. Open the Maintenance scheduler status screen and press
Perform
behind
the Clean System Bottles function.
2. Take out and empty the system wash bottle.
3. Clean with dilute hypochlorite (1 %) and rinse several times with water and
at the end with deionized water.
4. Refill with wash solution (for more information read the IFU of Wash Add
Ref 18971).
5. Take out and empty the waste bottle.
6. Clean with dilute hypochlorite (1 %) and rinse once with water.
7. Press
Ok
when finished, the scheduler will be updated.
8. Flush the System.
FIGURE 110