routIne taSKS
61
The window for automatic
Dilutions
will show the number of points that inte-
grate the curve. The
Serial factor
indicates the dilution factor: a factor of 2 will
define the concentration as 1/2, 1/4, 1/8, etc.; a factor of 3 will define concen-
trations as 1/3; 1/9; 1/27; etc. The set optionally can include the mother solu-
tion (1:1) and the blank. If included, they are within the defined number (i.e. Di-
lution # of 6 it is 1:1, 1:2, 1:4, 1:8, 1:16 and blank). The concentration of mother
solution is defined in the
Vial
sector. A minimum of 2 standards and a maximum
of 10 standards are accepted.
5.4.2 REQUESTING A CALIBRATION
Open the
Calibration
Menu and select on the left side the
Calibration
Submenu.
FIGURE 60
The instrument will automat-
ically adjust the number of
times this procedure is repeated
for completing all requested cali-
brators and replicates.