6
HT130003 / HT130042
HT130030 / HT130043
User Manual
CorTemp
®
Data Recorder v4.3
9001 Series
A.3. Turning ON the Data Recorder
NOTE: If the recorder has been OFF for a week or more, turn the recorder ON for
approximately 10 minutes. This allows the internal components to recharge. After
10 minutes, turn the recorder OFF. Then, after 1 minute, turn the recorder back ON
and initialize the Data Recorder. (Refer to Section III, Part A.12.: Initializing the Data
Recorder).
To turn
ON
the recorder, do the following:
1.
Once the battery has been installed turn the ON / OFF switch to the
ON
position.
2.
The recorder will run a self-test. During the self-test, screens similar to the following
will display. The Firmware Revision #, Heart Rate option, and System Frequency are
noted as below. The system frequency for HT130003 / HT130042 is 262kHz and no
system frequency is displayed on the recorder LCD, while the system frequency for
HT130030 / HT130043 is 300kHz and the system frequency is displayed on the
recorder LCD.
3.
On completion of the self-test, the date and time will appear briefly. The date
represents the last programmed date and the time represents the current time.
Verify
that the date is not 00-00-00 and that the time is correct.
A.4. Exit / Standby Mode
EXIT / STBY
mode allows the user to access the Standard Programming Menus, initiate
AUTO RUN MODE
, and
TURN OFF
the recorder.
CorTemp
®
CorTemp
®
CorTemp
®
CorTemp
®
CorTemp
®
Heart Rate Recorder
Non-Heart Rate Recorder
Heart Rate Designation
Firmware Revision #
System Frequency