17
HT130003 / HT130042
HT130030 / HT130043
User Manual
CorTemp
®
Data Recorder v4.3
9001 Series
beats and extrapolate heart rate per minute based on the time it took to sample the
programmed number of beats. It is recommended that the Heart Rate Interval is set to 3 to
ensure the greatest accuracy.
NOTE: In order to monitor heart rate, the Temperature Collection Interval must be
set at a minimum of 20 seconds.
To program the heart rate interval, do the following:
1. From
EXIT / STBY
mode, press
MENU
.
2. Press
READ
until the
H_RATE
/ HEART RATE ALARM LIMITS
screen displays.
3. Press
F1
to program the
H_RATE INTERVAL
.
4. Enter
0, 1
,
2
, or
3
.
NOTE: Default is 0 (OFF).
5. Press
ENTER
to return to
EXIT / STBY
mode.
A.22. High / Low Heart Rate Alarms
High and Low Heart Rate Alarms display on the recorder screen when a reading is outside
the programmed parameters. When using a Data Recorder Alarm, this reading will activate
an audio / visual alarm, while using the recorder in the
AUTO RUN MODE.
To program high / low heart rate alarms, do the following:
1. From
EXIT / STBY
mode, press
MENU
.
2. Press
READ
until the
H_RATE
/ HEART RATE ALARM LIMITS
screen displays.
3. Press
F2
to program the
HEART RATE ALARM LIMITS
.
4.
Enter the High Limit (HL) and the Low Limit (LL).
NOTE: Default is 000.0
5. Press
ENTER
to return to
EXIT / STBY
mode.
When the alarm limit is triggered,
♥
ALRM
will display on the recorder screen.
NOTE: If a temperature HALRM or LALRM is present, the
♥
ALRM will NOT display.
Heart Rate Reading below Low Limit
Temperature & Heart Rate Readings are above
respective High Limits
CorTemp
®
CorTemp
®
CorTemp
®
CorTemp
®