14
HT130003 / HT130042
HT130030 / HT130043
User Manual
CorTemp
®
Data Recorder v4.3
9001 Series
SENSOR TEST
can be used
to check sensors just received or alarming in storage due to magnet shifting,
to verify that a sensor is active before administering it to a subject (player),
to test whether or not a subject (player) may have passed the sensor from the
body, or
to detect signal interference (EMI), if no sensor is present.
To test a sensor, do the following:
1. From
EXIT / STBY
mode, press
MENU
.
2. Press
READ
until the
ID# / SENSOR TEST
screen displays.
3. Press
F2
to initiate
SENSOR TEST
.
4.
The recorder will continuously check for an active signal and will display SENSOR ON
or SENSOR OFF, depending on the actual status of the signal.
5. Press
ENTER
to return to the
ID# / SENSOR TEST
screen.
6. Press
ENTER
to return to
EXIT / STBY
mode.
A.18. Resetting LBAT
The
LBAT
error message will display on the recorder screen when any of the following
situations occur. The recorder has a
low
battery,
loses power, or
was turned off incorrectly.
Once detected, the recorder screen will continue to display
LBAT
until the
LBAT
is reset.
To reset the
LBAT
error message, do the following:
1. From
EXIT / STBY
mode, press
MENU
.
2. Press
READ
until the
SET BAT / DETECT SIG
screen displays.
CorTemp
®
CorTemp
®
CorTemp
®
CorTemp
®