15
HT130003 / HT130042
HT130030 / HT130043
User Manual
CorTemp
®
Data Recorder v4.3
9001 Series
3. Press
F1
to access
SET BAT
sub-menus. The screen will display
LBAT RESET
.
4. Press
F1
to reset
LBAT
. The screen will display
BAT OK
.
5. Press
ENTER
to return to the
SET BAT / DETECT SIG
screen.
6. Press
ENTER
to return to
EXIT / STBY
mode.
A.19. Low Battery Detector
The recorder can be set to detect low battery power and display a
LBAT
error message.
The recorder will end data collection when it detects low battery power and will go into
sleep mode, which disrupts extended monitoring periods. This feature may not be useful in
all situations. If the battery completely drains, the recorder will shutdown, without losing
data or programmed parameters. It is important not to let the recorder stay in this condition
for an extended period of time. If the internal components discharge completely, data and
parameters will be lost.
To set the recorder to detect low battery power, do the following:
1. From
EXIT / STBY
mode, press
MENU
.
2. Press
READ
until the
SET BAT / DETECT SIG
screen displays.
3. Press
F1
to access
SET BAT
sub-menus. The following screen will display:
4. Press
F2
to set the
DETECT LBAT
status,
DETECT LBAT
will toggle between
ON
and
OFF
.
NOTE: Default is ON.
5. Press
ENTER
to return to the
SET BAT / DETECT SIG
screen.
6. Press
ENTER
to return to
EXIT / STBY
mode.