Standard TV
The standard TV menu enables the oscilloscope to trigger on clamped TV
signals that use the 525 lines/60 Hz, 625 lines/50 Hz, 875 lines/60 Hz
standards.
Standard Softkey
The
standard
softkey allows you to select between
the three most common TV standards: 525 lines/60 Hz (NTSC) is the
standard used in the United States, 625 lines/50 Hz (PAL ) is the
standard used in most European and countries, and 875 lines/60 Hz is the
high-definition TV standard used by Zenith in the United States.
Field
The
field
softkey allows you to set the oscilloscope to trigger on
either field 1 and field 2 of the interlaced TV signal.
Line
The
line
softkey allows you to select which of the lines within the
chosen field you want the oscilloscope to trigger on. The trigger is
generated on the selected line. The range of lines you can choose from
depends on what the
standard
and
field
softkeys are set to.
When the
standard
softkey is set to 525 lines/60 Hz (which is compatible
with broadcast standard M), field 1 allows a range of lines from 1 to 263 and
field 2 allows a range of lines from 1 to 262. When the
standard
softkey is
set to 625 lines/50 Hz (which is compatible with broadcast standards B, C, D,
G, H, I, K, K1, L, and N), field 1 allows a range of lines from 1 to 313 and field
2 allows a range of lines from 314 to 625. When the
standard
softkey is set
to 875 lines/60 Hz, field 1 allows a range of lines is from 1 to 438 and field 2
allows a range of lines from 1 to 437.
Sync Polarity
The
Sync polarity
softkey allows you to trigger on
either a positive or negative sync pulse in the TV signal.
Trigger Menu
Mode
18–18
Содержание 54710A
Страница 7: ...vi ...
Страница 9: ...viii ...
Страница 16: ...Glossary Index Contents Contents 7 ...
Страница 17: ...Contents 8 ...
Страница 65: ...2 16 ...
Страница 72: ...Interpolation turned off Interpolation turned on Figure 3 2 Figure 3 3 Acquisition Menu Interpolate 3 7 ...
Страница 86: ...4 Applications ...
Страница 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Страница 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Страница 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Страница 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Страница 143: ...8 22 ...
Страница 144: ...9 Persistence 9 3 Color Grade 9 5 Draw waveform 9 6 Graticule 9 10 Label 9 14 Color 9 17 Display Menu ...
Страница 163: ...9 20 ...
Страница 164: ...10 Messages ...
Страница 170: ...11 Off 11 3 Manual 11 3 Waveform 11 5 Measurement 11 7 Marker Hints 11 8 Histogram 11 8 Marker Menu ...
Страница 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Страница 186: ...13 Measurements ...
Страница 199: ...Waveform definitions used to make voltage measurements Figure 13 10 Measurements Voltage Measurements 13 14 ...
Страница 239: ...Response of a Gaussian Filter Versus Frequency Figure 13 26 Measurements Time interval measurements 13 54 ...
Страница 245: ...13 60 ...
Страница 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Страница 261: ...15 10 ...
Страница 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Страница 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Страница 319: ...19 18 ...
Страница 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Страница 328: ...21 Display 21 3 Source 21 3 Window 21 3 FFT Scaling 21 4 FFTs and Automatic Measurements 21 8 FFT Basics 21 10 FFT Menu ...
Страница 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Страница 357: ...22 14 ...
Страница 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Страница 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Страница 407: ...Index 10 ...
Страница 409: ......