Memory error occurred in plug-in ?: Try reinstalling plug-in
The checksum or contents of the plug-in’s memory is incorrect. The ?
indicates which plug-in has the error. Check the following:
Reinstall the plug-in.
Finger tighten the knurled screw on the front-panel of the plug-in.
If these actions do not solve the problem, there must be a hardware problem
in the mainframe or plug-in. Call Hewlett-Packard for service.
Overload condition on ?
The input signal on channel ? is too big. The overload protection hardware
has been activated. The message stays on screen until the overload is
removed. Check the appropriate plug-in manual for overload limits. Each
plug-in is different, and not all have overload protection.
Please correct overload condition, then press Overload Cleared
This is further prompting that occurs when the menu of an overloaded
channel is selected. Actions to take include the following:
•
Removing the input signal
•
Reducing the amplitude of the input signal
•
Decreasing the channel sensitivity
•
Changing the input coupling from dc to ac (if the signal has a large offset)
Plug-in configuration has changed
The physical configuration of the plug-ins is different from when the setup
was saved to memory or disk. The system has tried to establish the same
setup.
Plug-in ? is not supported: System firmware upgrade is needed
In the future, new plug-ins may require a certain revision of system
firmware for proper operation. The ? indicates which plug-in is not
supported. Use the upgrade service to keep the system firmware current.
Messages
10-5
Содержание 54710A
Страница 7: ...vi ...
Страница 9: ...viii ...
Страница 16: ...Glossary Index Contents Contents 7 ...
Страница 17: ...Contents 8 ...
Страница 65: ...2 16 ...
Страница 72: ...Interpolation turned off Interpolation turned on Figure 3 2 Figure 3 3 Acquisition Menu Interpolate 3 7 ...
Страница 86: ...4 Applications ...
Страница 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Страница 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Страница 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Страница 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Страница 143: ...8 22 ...
Страница 144: ...9 Persistence 9 3 Color Grade 9 5 Draw waveform 9 6 Graticule 9 10 Label 9 14 Color 9 17 Display Menu ...
Страница 163: ...9 20 ...
Страница 164: ...10 Messages ...
Страница 170: ...11 Off 11 3 Manual 11 3 Waveform 11 5 Measurement 11 7 Marker Hints 11 8 Histogram 11 8 Marker Menu ...
Страница 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Страница 186: ...13 Measurements ...
Страница 199: ...Waveform definitions used to make voltage measurements Figure 13 10 Measurements Voltage Measurements 13 14 ...
Страница 239: ...Response of a Gaussian Filter Versus Frequency Figure 13 26 Measurements Time interval measurements 13 54 ...
Страница 245: ...13 60 ...
Страница 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Страница 261: ...15 10 ...
Страница 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Страница 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Страница 319: ...19 18 ...
Страница 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Страница 328: ...21 Display 21 3 Source 21 3 Window 21 3 FFT Scaling 21 4 FFTs and Automatic Measurements 21 8 FFT Basics 21 10 FFT Menu ...
Страница 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Страница 357: ...22 14 ...
Страница 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Страница 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Страница 407: ...Index 10 ...
Страница 409: ......