When a single-wide plug-in and a two-wide plug-in are both turned on, both
plug-ins sample at the same sample rate up to 2 GSa/s. Above 2 GSa/s, the
single-wide plug-in samples at its maximum sample rate of 2 GSa/s while the
two-wide plug-in samples at the selected sample rate above 2 GSa/s.
Remember, even though you can select a 4-GSa/s sample rate or an 8-GSa/s
sample rate, a single-wide plug-in has a maximum sample rate of 2 GSa/s.
If the two-wide is turned off and the single-wide is turned on, the status line
displays the actual sample rate up to 2 GSa/s. However, the acquisition menu
allows you to select a sample rate up to 4 GSa/s or up to 8 GSa/s. That is
because even though the two-wide plug-in or the four-wide plug-in is turned
off, if it is used as an operand in a waveform math function, it is acquiring
data at the selected sample rate.
Automatic
Automatic lets the oscilloscope select the sampling rate for you. The
advantage is that the oscilloscope selects a sample rate that optimizes the
way the waveform is displayed and the display update rate.
When the record length is set to manual and sample rate is set to automatic,
the oscilloscope selects the highest sample rate that allows all the data in the
waveform record to fill the screen.
Manual
Manual lets you pick a sample rate. However, you have to make the trade-off
decisions on the interaction among sample rate, record length, and
measurement throughput.
Acquisition Menu
Sampling Rate
3–13
Содержание 54710A
Страница 7: ...vi ...
Страница 9: ...viii ...
Страница 16: ...Glossary Index Contents Contents 7 ...
Страница 17: ...Contents 8 ...
Страница 65: ...2 16 ...
Страница 72: ...Interpolation turned off Interpolation turned on Figure 3 2 Figure 3 3 Acquisition Menu Interpolate 3 7 ...
Страница 86: ...4 Applications ...
Страница 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Страница 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Страница 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Страница 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Страница 143: ...8 22 ...
Страница 144: ...9 Persistence 9 3 Color Grade 9 5 Draw waveform 9 6 Graticule 9 10 Label 9 14 Color 9 17 Display Menu ...
Страница 163: ...9 20 ...
Страница 164: ...10 Messages ...
Страница 170: ...11 Off 11 3 Manual 11 3 Waveform 11 5 Measurement 11 7 Marker Hints 11 8 Histogram 11 8 Marker Menu ...
Страница 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Страница 186: ...13 Measurements ...
Страница 199: ...Waveform definitions used to make voltage measurements Figure 13 10 Measurements Voltage Measurements 13 14 ...
Страница 239: ...Response of a Gaussian Filter Versus Frequency Figure 13 26 Measurements Time interval measurements 13 54 ...
Страница 245: ...13 60 ...
Страница 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Страница 261: ...15 10 ...
Страница 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Страница 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Страница 319: ...19 18 ...
Страница 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Страница 328: ...21 Display 21 3 Source 21 3 Window 21 3 FFT Scaling 21 4 FFTs and Automatic Measurements 21 8 FFT Basics 21 10 FFT Menu ...
Страница 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Страница 357: ...22 14 ...
Страница 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Страница 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Страница 407: ...Index 10 ...
Страница 409: ......