Delay events
Delay by events has the oscilloscope arm on an edge from one of the channel
or trigger inputs, wait for a number of events that you specify in the menu,
then trigger on an edge from any of the other channel or trigger inputs.
Basically you can think of delay by events as two edge triggers that are
separated by a selectable number of events. You can arm the trigger circuit
on one slot, count events on a second slot, then trigger on a third slot.
You can use delay by events for reducing the jitter on long delay
measurements on a pulse burst or pulse train. If you use a long time base
delay to look at a particular pulse in a pulse burst or pulse train, the jitter on
the pulse can make it almost impossible to see the desired pulse. Delay by
events allows you to delay out to the pulse of interest, then trigger on that
pulse. Because you are triggering on the pulse of interest, jitter is not a
factor in viewing the pulse.
For example, you may want to look at the 97th pulse out of 100 pulses. You
can arm on the first pulse, then delay out 95 events, then trigger on the 97th
pulse. By varying the number of events, you can consecutively look at pulses
before and after the pulse of interest.
Arm on
Selects the slot, trigger level, and slope for the arming
conditions of the trigger circuitry.
Event
Selects the slot, trigger level, and slope for the event the trigger
circuitry is to count.
Delay
Used to set the number of events to count from 0 to 16 million.
Trigger
Selects the slot, trigger level, and slope for the trigger event.
Trigger Menu
Mode
18–16
Содержание 54710A
Страница 7: ...vi ...
Страница 9: ...viii ...
Страница 16: ...Glossary Index Contents Contents 7 ...
Страница 17: ...Contents 8 ...
Страница 65: ...2 16 ...
Страница 72: ...Interpolation turned off Interpolation turned on Figure 3 2 Figure 3 3 Acquisition Menu Interpolate 3 7 ...
Страница 86: ...4 Applications ...
Страница 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Страница 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Страница 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Страница 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Страница 143: ...8 22 ...
Страница 144: ...9 Persistence 9 3 Color Grade 9 5 Draw waveform 9 6 Graticule 9 10 Label 9 14 Color 9 17 Display Menu ...
Страница 163: ...9 20 ...
Страница 164: ...10 Messages ...
Страница 170: ...11 Off 11 3 Manual 11 3 Waveform 11 5 Measurement 11 7 Marker Hints 11 8 Histogram 11 8 Marker Menu ...
Страница 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Страница 186: ...13 Measurements ...
Страница 199: ...Waveform definitions used to make voltage measurements Figure 13 10 Measurements Voltage Measurements 13 14 ...
Страница 239: ...Response of a Gaussian Filter Versus Frequency Figure 13 26 Measurements Time interval measurements 13 54 ...
Страница 245: ...13 60 ...
Страница 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Страница 261: ...15 10 ...
Страница 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Страница 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Страница 319: ...19 18 ...
Страница 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Страница 328: ...21 Display 21 3 Source 21 3 Window 21 3 FFT Scaling 21 4 FFTs and Automatic Measurements 21 8 FFT Basics 21 10 FFT Menu ...
Страница 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Страница 357: ...22 14 ...
Страница 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Страница 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Страница 407: ...Index 10 ...
Страница 409: ......