Normal Accuracy Calibration Level
Normal accuracy calibration level refers to the level of accuracy that is
achieved when a plug-in is installed in a slot of a mainframe, but has not been
calibrated to best accuracy in that particular slot.
The oscilloscope uses the mainframe calibration factors and the plug-in
calibration factors to determine the normal calibration factors for that plug-in
in that slot.
The intent of this level of calibration is to allow any plug-in to be inserted in
any mainframe slot and still achieve a typical vertical accuracy of
±
3%.
For two-wide plug-ins, like the HP 54721A, the interleaving of the two
mainframe slots is performed during the best accuracy calibration.
Therefore, the real-time sampling mode is typically not useful in the normal
accuracy mode. For accurate results, a best accuracy calibration must be
performed to use two-wide plug-ins in the real-time mode.
See Also
"Best Accuracy" in this chapter for details on the best accuracy calibration.
"Normal Accuracy" in chapter 16 for more details on the normal accuracy
specifications.
Calibration Overview
Normal Accuracy Calibration Level
5–5
Содержание 54710A
Страница 7: ...vi ...
Страница 9: ...viii ...
Страница 16: ...Glossary Index Contents Contents 7 ...
Страница 17: ...Contents 8 ...
Страница 65: ...2 16 ...
Страница 72: ...Interpolation turned off Interpolation turned on Figure 3 2 Figure 3 3 Acquisition Menu Interpolate 3 7 ...
Страница 86: ...4 Applications ...
Страница 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Страница 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Страница 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Страница 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Страница 143: ...8 22 ...
Страница 144: ...9 Persistence 9 3 Color Grade 9 5 Draw waveform 9 6 Graticule 9 10 Label 9 14 Color 9 17 Display Menu ...
Страница 163: ...9 20 ...
Страница 164: ...10 Messages ...
Страница 170: ...11 Off 11 3 Manual 11 3 Waveform 11 5 Measurement 11 7 Marker Hints 11 8 Histogram 11 8 Marker Menu ...
Страница 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Страница 186: ...13 Measurements ...
Страница 199: ...Waveform definitions used to make voltage measurements Figure 13 10 Measurements Voltage Measurements 13 14 ...
Страница 239: ...Response of a Gaussian Filter Versus Frequency Figure 13 26 Measurements Time interval measurements 13 54 ...
Страница 245: ...13 60 ...
Страница 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Страница 261: ...15 10 ...
Страница 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Страница 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Страница 319: ...19 18 ...
Страница 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Страница 328: ...21 Display 21 3 Source 21 3 Window 21 3 FFT Scaling 21 4 FFTs and Automatic Measurements 21 8 FFT Basics 21 10 FFT Menu ...
Страница 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Страница 357: ...22 14 ...
Страница 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Страница 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Страница 407: ...Index 10 ...
Страница 409: ......