Control setting is unchanged: Entry was not completed
The enter key was not pressed after typing in numbers using the keypad
before changing menus. If the entry still desired, it will have to be completely
entered again.
Execution not possible: ? has no acquisition capability
The plug-in indicated by the ? does not have acquisition capability. Select
another plug-in that does have acquisition capability. For example, the HP
54718A plug-in only has triggering capability.
Function cannot be performed on the selected waveform
The function is not defined for this waveform type, and, thus, cannot be
performed.
Id error occurred in plug-in ?: Service is required
The information the mainframe downloaded from the plug-in’s memory is
inconsistent with the plug-in’s hardware. The ? indicates which plug-in has
the Id error. This is a special case of memory error. Check the following:
Reinstall the plug-in.
Finger tighten the knurled screw on the front-panel of the plug-in.
If these actions do not solve the problem, there must be a hardware problem
in mainframe or plug-in. Call Hewlett-Packard service.
Incompatible setup
It is possible that due to a system firmware change, a previously saved setup
may become incompatible.
Low battery voltage detected: Service is required
Change the battery as soon as this message is displayed. Cal factors and data
may be lost when the power is cycled, and the instrument may not function
correctly. When Hewlett-Packard service replaces the battery, a revised boot
ROM may be installed.
Measurement cannot be performed on the selected waveform
The measurement is not defined for this waveform type, and cannot be made.
For example, the rise time of an A versus B waveform is not defined.
Messages
10-4
Содержание 54710A
Страница 7: ...vi ...
Страница 9: ...viii ...
Страница 16: ...Glossary Index Contents Contents 7 ...
Страница 17: ...Contents 8 ...
Страница 65: ...2 16 ...
Страница 72: ...Interpolation turned off Interpolation turned on Figure 3 2 Figure 3 3 Acquisition Menu Interpolate 3 7 ...
Страница 86: ...4 Applications ...
Страница 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Страница 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Страница 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Страница 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Страница 143: ...8 22 ...
Страница 144: ...9 Persistence 9 3 Color Grade 9 5 Draw waveform 9 6 Graticule 9 10 Label 9 14 Color 9 17 Display Menu ...
Страница 163: ...9 20 ...
Страница 164: ...10 Messages ...
Страница 170: ...11 Off 11 3 Manual 11 3 Waveform 11 5 Measurement 11 7 Marker Hints 11 8 Histogram 11 8 Marker Menu ...
Страница 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Страница 186: ...13 Measurements ...
Страница 199: ...Waveform definitions used to make voltage measurements Figure 13 10 Measurements Voltage Measurements 13 14 ...
Страница 239: ...Response of a Gaussian Filter Versus Frequency Figure 13 26 Measurements Time interval measurements 13 54 ...
Страница 245: ...13 60 ...
Страница 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Страница 261: ...15 10 ...
Страница 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Страница 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Страница 319: ...19 18 ...
Страница 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Страница 328: ...21 Display 21 3 Source 21 3 Window 21 3 FFT Scaling 21 4 FFTs and Automatic Measurements 21 8 FFT Basics 21 10 FFT Menu ...
Страница 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Страница 357: ...22 14 ...
Страница 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Страница 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Страница 407: ...Index 10 ...
Страница 409: ......