The digitizer samples the applied signal and converts it to a digital signal.
The FISO holds the data until the system bus is ready for the data. The
output of the FISO is used as an address to the calibration read-through table
(cal table). The cal table automatically applies the calibration factors to the
data.
In the real-time sampling mode, the calibrated data is stored in the channel
memories before any of the postprocessing is performed. Postprocessing
includes turning on or off the digital bandwidth limit filter or the interpolator,
calculating functions, storing data to the waveform memories, transferring
data over the HP-IB bus, or transferring data to and from the disk. Notice
that the measurements are performed on the real-time data after it has gone
through postprocessing.
Therefore, you can make measurements on the data, and you can turn on or
off digital bandwidth limit or interpolation without having to reacquire the
data. This is important because the real-time sampling mode is primarily
used on events that happen either once or infrequently, and reacquiring the
data may not be one of your options. Also, turning on interpolation usually
improves the repeatability of your measurements.
The equivalent-time sampling mode is slightly different. Notice that
averaging is turned on or off before the data is stored in the channel
memories. That means once the data is acquired, if you need to turn
averaging on or off before making any measurements, you must reacquire the
data. However, because the equivalent-time sampling mode is primarily used
on repetitive signals, you should be able to reacquire the data.
How the Oscilloscope Works
Data Flow
1–9
Содержание 54710A
Страница 7: ...vi ...
Страница 9: ...viii ...
Страница 16: ...Glossary Index Contents Contents 7 ...
Страница 17: ...Contents 8 ...
Страница 65: ...2 16 ...
Страница 72: ...Interpolation turned off Interpolation turned on Figure 3 2 Figure 3 3 Acquisition Menu Interpolate 3 7 ...
Страница 86: ...4 Applications ...
Страница 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Страница 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Страница 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Страница 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Страница 143: ...8 22 ...
Страница 144: ...9 Persistence 9 3 Color Grade 9 5 Draw waveform 9 6 Graticule 9 10 Label 9 14 Color 9 17 Display Menu ...
Страница 163: ...9 20 ...
Страница 164: ...10 Messages ...
Страница 170: ...11 Off 11 3 Manual 11 3 Waveform 11 5 Measurement 11 7 Marker Hints 11 8 Histogram 11 8 Marker Menu ...
Страница 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Страница 186: ...13 Measurements ...
Страница 199: ...Waveform definitions used to make voltage measurements Figure 13 10 Measurements Voltage Measurements 13 14 ...
Страница 239: ...Response of a Gaussian Filter Versus Frequency Figure 13 26 Measurements Time interval measurements 13 54 ...
Страница 245: ...13 60 ...
Страница 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Страница 261: ...15 10 ...
Страница 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Страница 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Страница 319: ...19 18 ...
Страница 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Страница 328: ...21 Display 21 3 Source 21 3 Window 21 3 FFT Scaling 21 4 FFTs and Automatic Measurements 21 8 FFT Basics 21 10 FFT Menu ...
Страница 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Страница 357: ...22 14 ...
Страница 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Страница 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Страница 407: ...Index 10 ...
Страница 409: ......