To use the external trigger input
1
Connect a signal to the BNC labeled “Ext trigger in.”
The “Ext trigger in” BNC is the rightmost one on the HP 54620A/C front
panel. It has an input impedance of 1 M
Ω
and can accept a maximum signal
of
±
40 V. You can connect the trigger output of other instruments to the
external trigger input, or use an oscilloscope probe, such as the HP 10071A.
2
Set up the proper threshold voltage for the input using the Logic
levels menu.
See “To change the logic threshold for input signals.” The external trigger
threshold is represented as
E Trg In
under the
Set
softkey.
3
Set up an edge, pattern, or advanced trigger specification using the
external trigger input as part of the specification.
The external trigger input is labeled as “Ext” in the pattern, and is selectable
with either the Select knob or the Entry knob during pattern or edge
definition. See “To define an edge trigger,” “To define a pattern trigger,” or
“To define an advanced trigger.”
The external trigger input is useful when it is not practical to completely
qualify the desired trigger pattern using only the logic inputs. For example,
you may want to trigger the analyzer only after another instrument has found
a trigger. Or, you may need to use all logic channels for input data, and need
just one additional channel for an edge trigger. Or, you may want to use a
scope probe to examine a system signal with larger signal swings than
allowed by the logic inputs.
Making Analyzer Measurements
To use the external trigger input
108
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...