![HP 54620A Скачать руководство пользователя страница 80](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475080.webp)
All Qualifiers must be Satisfied to Trigger the Analyzer
When setting up an advanced trigger, you must remember that all required
events must be satisfied to trigger the analyzer. For example, if you set up an
occurrence trigger using Pattern 1 and Edge 1, and the occurrence count is 3,
then Edge 1 must be satisfied three times, with the pattern valid each time,
before the trigger will occur.
Pattern 1 and Edge 1 are Shared Sources
The source definitions for Pattern 1 and Edge 1 are shared with the pattern
trigger specification. Thus, changes here will affect that specification, and
changes to the simple pattern trigger specification will affect the definitions of
Pattern 1 and Edge 1. Defining patterns for a pattern trigger is different. See “To
define a pattern trigger.”
Making Analyzer Measurements
To define an advanced trigger
79
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...