![HP 54620A Скачать руководство пользователя страница 105](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475105.webp)
Using the Analyzer with Other Instruments
The HP 54620A/C provides two features that help extend the
triggering capabilities of the analyzer and allow you to use it with
other instruments:
•
External trigger input
•
Trigger output
Extending trigger capabilities
External Trigger Input
The external trigger input allows you to trigger the HP 54620A/C from
another instrument or another signal in the circuit under test.
•
In the first case, you may be using the capability of the other instrument to
help qualify the trigger condition before triggering the analyzer.
•
In the second case, you may simply need a “17th channel” or need to
trigger from an analog signal where the inputs are not directly appropriate
for analyzer input, but which can be input safely using a 10:1 divider probe.
The external trigger signal is available in edge, pattern, and advanced trigger
to use as either an edge or pattern qualifier for the trigger condition.
104
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...