Figure 22 shows the same acquisition with the horizontal mode set to
Delayed. Again, the time reference is set to Center. Though the marker (the
solid triangle
▼
) points to the trigger event in both the main and delayed
sweep, the delay values themselves are independently adjustable. The delay
value for the delayed sweep is set to 8
µ
s. This is different than the setting
for the main sweep, which remains at 15
µ
s in this figure, and is not
adjustable with the delay knob (while the delayed sweep is active). In both
the main and delayed sweep, the hollowed triangle (
∇
) indicates the part of
the waveform that is delayed by the respective value from the trigger event.
Delayed Sweep with Delay of 8
µ
s
Empty triangle in main sweep
shows the part of the
waveform that is delayed
15
µ
s from the trigger
The trigger position in
main and delayed
sweeps
Empty triangle in delayed
sweep shows the part of
the waveform that is
delayed 8
µ
s from the
trigger
Figure 22
Making Analyzer Measurements
To change the time reference position
85
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...