Table 19
Acquisition System
Maximum Sample Rate
500 MSa/s.
Resolution
Single bit.
Simultaneous Channels
16.
Record Length
2K Samples at sampling periods of 8 ns and
slower ( sweep speeds of 1
µ
S/div to 1
S/div).
8K samples at sampling periods of 2 ns and
4 ns (sweep speeds of 5 nS/div to
500 nS/div).
When Auto Glitch Detect is disabled, its
acquisition system has a record length of 8K
samples at all sweep speeds. When Auto
Glitch Detect is disabled, you can increase
the sweep speed to 2 s/div and 5 s/div in the
Main display mode.
Minimum Update Rate
10 full screens per second independent of
the number of channels being displayed
(assumes no measurements).
Glitch Capture
Automatically activated at all sweep speeds
where sampling period is slowed to be
greater than 4 ns (1
µ
S/div and slower)
when Auto Glitch Detect is enabled. Will
capture pulses as narrow as 3.5 ns at all
activated sweep speeds. Maximum glitch
width is sample period - 1 ns.
Performance Characteristics
199
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...