![HP 54620A Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475056.webp)
Controlling Data Acquisition
During an
acquisition
, the logic analyzer examines the input voltage
at each input probe at a sampling rate defined by the time base setting
(Time/Div knob). At each sample, it compares the input voltage to the
logic threshold. If the voltage is above the threshold, the analyzer
stores a “1” in sample memory; otherwise, it stores a “0.” When sample
memory is full, the display logic reconstructs the input waveform from
the pattern of bits stored in sample memory. To control the
HP 54620A/C acquisition process, you can
•
perform continuous acquisitions by pressing the Run/Stop key, and
stop them by pressing it again,
•
make one acquisition, then stop, by pressing the Single key,
•
store the results of several acquisitions by pressing Autostore, or
•
erase the results of all acquisitions by pressing Erase.
Auto Glitch Detect
The HP 54620A/C automatically enables Auto Glitch Detect upon
powerup. When the Auto Glitch Detect feature is enabled, it puts the
instrument in glitch detect mode at all sweep speeds of 1
µ
s/div and
slower where the sampling speed must be reduced, thereby
preventing aliases from being displayed.
When Auto Glitch Detect is disabled, the HP 54620A/C acquisition
system has a record length of 8K samples at all sweep speeds. In
addition, you can extend the sweep speed to 2 s/div and 5 s/div in the
Main display mode. See “Glitch Detection” in chapter 4.
When Auto Glitch Detect is enabled, a waveform display that contains
glitches, separated by the sampling intervals, is an indication that the
waveform is being undersampled. A faster Time/Div setting is needed
to obtain a correct display of the waveform. This easy indication of
undersampling is not available when Auto Glitch Detect is disabled.
55
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...