![HP 54620A Скачать руководство пользователя страница 202](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475202.webp)
External
trigger
Input R & C
~ 1 Meg
Ω
and 12 pF, compatible with HP 1007X
probes.
Maximum Input
±
40 V peak.
Trigger Threshold
±
6 volts, settable in 50-mV increments.
Threshold Accuracy
1
±
100 mV or 6% of setting, whichever is greater.
Minimum Input Change
200 mV pp.
Minimum Pulse Width
20 ns.
Trigger
Output
Output is a rising edge at the trigger point.
Output Level
0 to
≥
2.0V into 50
Ω
0 to
≥
4.8V open circuit.
Delay, data in to
trigger out
~ 85 ns.
Jitter
±
Sample 10 ppm.
Maximum Output Rate
≅
2 kHz (HP 54620A),
≅
1.5 kHz (HP 54620C) with the
analyzer stopped, 20/sec running.
Performance Characteristics
201