![HP 54620A Скачать руководство пользователя страница 55](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475055.webp)
To change the logic threshold for input signals
1
Press
Logic Levels
.
2
Press the
Set
softkey to highlight the range of channels for which
you want to set the logic threshold.
Ranges available are
Chan 0-7
(channels 0 through 7),
Chan 8-15
(channels 8 through 15), and
TRIG IN
(the external trigger input).
3
Press
TTL
,
CMOS
,
ECL
, or
User
to select the threshold.
4
If you selected
User
, set the threshold voltage using the Entry knob.
Turning the Entry knob while in this menu automatically sets the Threshold
option to
User
.
5
Repeat steps 2 through 4 for each group of channels for which you
want to set the threshold.
You can exit the threshold setting menu by pressing any other front-panel
key.
The threshold voltage setting is used by the input comparators to determine
whether an input signal is a logic low or logic high. The settings for each
option are shown in table 2.
Table 3
Threshold Voltage Settings
Option
Voltage
TTL
1.50 V
CMOS
2.50 V
ECL
-1.30 V
User
Continuously variable from –6.00 V to +6.00 V
Making Analyzer Measurements
To change the logic threshold for input signals
54
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...