3
Optimize the Time Null of the Self Cal.
a
Connect two BNC 50-
Ω
, 9-inch cables to the BNC tee adapter.
Connect the BNC 50
Ω
(f)(f) adapter to the BNC tee adapter, and
connect the 48-inch BNC cable to the BNC 50
Ω
(f)(f) adapter. Once
you select Start, the instrument will prompt you to connect the cables
to the appropriate locations on the rear panel of the instrument.
b
Select the Procedure field, then select Time Null from the pop-up
menu.
c
Select the Start field and follow the instructions on the display.
d
After completion of the Time Null calibration, remove the cables from
the instrument.
4
Calibrate the Logic Trigger of the Self Cal.
a
Obtain a BNC 50-
Ω
, 48-inch cable.
b
Select Start. The instrument will prompt you to connect the cable to
the appropriate location on the rear panel of the instrument.
c
Select the Procedure field, then select Logic Trigger from the pop-up
menu.
d
Select the Start field and follow the instructions on the display.
e
After completion of the Logic Trigger calibration, remove the cable
from the instrument.
Calibrating the oscilloscope
Self Cal menu calibrations
12-10
Содержание 1660CS
Страница 5: ...Introduction iv ...
Страница 16: ...1 Logic Analyzer Overview ...
Страница 24: ...2 Connecting Peripherals ...
Страница 33: ...2 10 ...
Страница 34: ...3 Using the Logic Analyzer ...
Страница 55: ...3 22 ...
Страница 56: ...4 Using the Trigger Menu ...
Страница 75: ...4 20 ...
Страница 76: ...5 Triggering Examples ...
Страница 109: ...5 34 ...
Страница 110: ...6 File Management ...
Страница 114: ...Saving the system configuration for programmatic control File Management To save a configuration 6 5 ...
Страница 119: ...6 10 ...
Страница 120: ...7 Reference ...
Страница 221: ...7 102 ...
Страница 222: ...8 System Performance Analysis SPA Software ...
Страница 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Страница 252: ...9 Concepts ...
Страница 273: ...HP 1660CS series analyzer theory HP 1660C CS logic analyzer The Analyzer Hardware HP 1660CS series analyzer theory 9 22 ...
Страница 279: ...Oscilloscope board theory Oscilloscope board The Analyzer Hardware Oscilloscope board theory 9 28 ...
Страница 284: ...10 Troubleshooting ...
Страница 298: ...11 Specifications ...
Страница 311: ...11 14 ...
Страница 312: ...12 Operator s Service ...
Страница 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Страница 337: ...Glossary 4 ...