To set up time correlation between two state analyzers
To correlate the data between two state analyzers, both must have Count
Time turned on in their Trigger menus. Although both have Count State
available, it is not possible to correlate data based on states even when they
are identically defined.
1
In the Analyzer Trigger menu, select Count.
Count may be Count Off, Count Time, or Count States. Selecting the field
causes a pop-up to appear.
2
Select the field after Count: and select Time.
A warning may appear about reduced memory. It will not prevent you from
changing Count to Count Time.
3
Select Done.
4
Repeat steps 1 through 3 for the other state analyzer.
Now when you acquire data you will be able to interleave the listings.
To set up time correlation between a timing and a
state analyzer
To set up time correlation between a timing and a state analyzer, only the
state analyzer needs to have Count Time turned on. The timing analyzer
automatically keeps track of time.
1
In the state Analyzer Trigger menu, select Count.
Count may be Count Off, Count Time, or Count States. Selecting the field
causes a pop-up to appear.
2
Select the field after Count: and select Time.
A warning may appear about reduced memory. It will not prevent you from
changing Count to Count Time.
3
Select Done.
Now when you acquire data you will be able to set up a mixed display.
Using the Trigger Menu
To set up time correlation between two state analyzers
4–11
Содержание 1660CS
Страница 5: ...Introduction iv ...
Страница 16: ...1 Logic Analyzer Overview ...
Страница 24: ...2 Connecting Peripherals ...
Страница 33: ...2 10 ...
Страница 34: ...3 Using the Logic Analyzer ...
Страница 55: ...3 22 ...
Страница 56: ...4 Using the Trigger Menu ...
Страница 75: ...4 20 ...
Страница 76: ...5 Triggering Examples ...
Страница 109: ...5 34 ...
Страница 110: ...6 File Management ...
Страница 114: ...Saving the system configuration for programmatic control File Management To save a configuration 6 5 ...
Страница 119: ...6 10 ...
Страница 120: ...7 Reference ...
Страница 221: ...7 102 ...
Страница 222: ...8 System Performance Analysis SPA Software ...
Страница 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Страница 252: ...9 Concepts ...
Страница 273: ...HP 1660CS series analyzer theory HP 1660C CS logic analyzer The Analyzer Hardware HP 1660CS series analyzer theory 9 22 ...
Страница 279: ...Oscilloscope board theory Oscilloscope board The Analyzer Hardware Oscilloscope board theory 9 28 ...
Страница 284: ...10 Troubleshooting ...
Страница 298: ...11 Specifications ...
Страница 311: ...11 14 ...
Страница 312: ...12 Operator s Service ...
Страница 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Страница 337: ...Glossary 4 ...