To place the trigger in memory
In Automatic Acquisition Mode, the exact location of the trigger depends on
the trigger specification but usually falls around the center. You can
manually place it to be at the beginning, end, or anywhere else.
1
In the Analyzer Trigger menu, select Acquisition Control.
The Acquisition Control menu pops up. If the acquisition mode is set to
Automatic, the menu contains a single field and an explanation. If Acquisition
has been customized, it has 3 or 4 fields and a picture showing where the
trigger is currently placed in memory.
2
If the mode is Automatic, select the field to toggle it to Manual.
The menu now shows three fields and a picture.
3
Select the Trigger Position field.
A list of choices appears, as shown in the timing trigger example below.
4
Select the appropriate entry for your needs.
Start, Center, and End place the trigger respectively at the beginning, middle,
and end of the memory. Delay, available in timing analyzers only, causes the
analyzer to not save any data before the time delay has elapsed. User Defined
calls up a fourth field where you specify exactly where you want the trigger.
5
Select Done.
Acquisition Control menu with trigger position pop-up for a timing analyzer
Using the Trigger Menu
To place the trigger in memory
4–18
Содержание 1660CS
Страница 5: ...Introduction iv ...
Страница 16: ...1 Logic Analyzer Overview ...
Страница 24: ...2 Connecting Peripherals ...
Страница 33: ...2 10 ...
Страница 34: ...3 Using the Logic Analyzer ...
Страница 55: ...3 22 ...
Страница 56: ...4 Using the Trigger Menu ...
Страница 75: ...4 20 ...
Страница 76: ...5 Triggering Examples ...
Страница 109: ...5 34 ...
Страница 110: ...6 File Management ...
Страница 114: ...Saving the system configuration for programmatic control File Management To save a configuration 6 5 ...
Страница 119: ...6 10 ...
Страница 120: ...7 Reference ...
Страница 221: ...7 102 ...
Страница 222: ...8 System Performance Analysis SPA Software ...
Страница 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Страница 252: ...9 Concepts ...
Страница 273: ...HP 1660CS series analyzer theory HP 1660C CS logic analyzer The Analyzer Hardware HP 1660CS series analyzer theory 9 22 ...
Страница 279: ...Oscilloscope board theory Oscilloscope board The Analyzer Hardware Oscilloscope board theory 9 28 ...
Страница 284: ...10 Troubleshooting ...
Страница 298: ...11 Specifications ...
Страница 311: ...11 14 ...
Страница 312: ...12 Operator s Service ...
Страница 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Страница 337: ...Glossary 4 ...