Using State Histogram mode
Choosing a label to monitor
To specify a label to monitor, select the
Label field in the State Histogram display (item 1 in the figure below). In
the pop-up, you will see a list of all the labels defined in the Format
specification. From this list, choose the label you want to monitor.
Changing from one label to another will erase the display setup for the first
label. If you want to change to a different label, but do not want to lose the
setup for the current one, save the current one to a disk.
Defining the ranges
To define a range on the specified label, select
one of the 11 range fields (item 2 in the figure below). When you select
one of the 11 ranges, you will see the Range Definition pop-up (item 3 in
the figure below). This pop-up has fields where you enter the low and
high value, a name for the range, and whether the range is on or off.
SPA State Histogram menu with fields called out
2
3
1
System Performance Analysis (SPA) Software
Using State Overview, State Histogram, and Time Interval
8-24
Содержание 1660CS
Страница 5: ...Introduction iv ...
Страница 16: ...1 Logic Analyzer Overview ...
Страница 24: ...2 Connecting Peripherals ...
Страница 33: ...2 10 ...
Страница 34: ...3 Using the Logic Analyzer ...
Страница 55: ...3 22 ...
Страница 56: ...4 Using the Trigger Menu ...
Страница 75: ...4 20 ...
Страница 76: ...5 Triggering Examples ...
Страница 109: ...5 34 ...
Страница 110: ...6 File Management ...
Страница 114: ...Saving the system configuration for programmatic control File Management To save a configuration 6 5 ...
Страница 119: ...6 10 ...
Страница 120: ...7 Reference ...
Страница 221: ...7 102 ...
Страница 222: ...8 System Performance Analysis SPA Software ...
Страница 241: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 8 20 ...
Страница 252: ...9 Concepts ...
Страница 273: ...HP 1660CS series analyzer theory HP 1660C CS logic analyzer The Analyzer Hardware HP 1660CS series analyzer theory 9 22 ...
Страница 279: ...Oscilloscope board theory Oscilloscope board The Analyzer Hardware Oscilloscope board theory 9 28 ...
Страница 284: ...10 Troubleshooting ...
Страница 298: ...11 Specifications ...
Страница 311: ...11 14 ...
Страница 312: ...12 Operator s Service ...
Страница 324: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 12 13 ...
Страница 337: ...Glossary 4 ...