Data
67
J
Zero filter
J
Automatic calibration unit
J
Manual calibration unit
Product name
Zero filter unit
Model
RA-ZF
Operating range
0.1 L/min to 0.5 L/min
Material of liquid contact section
PVC, EPDM, PMMA
Operating temperature
0
°
C to 50
°
C (no freezing)
Primary pressure
0.1 MPa to 0.2 MPa
Piping connection
sample inlet and outlet
φ
8 mm
×
φ
6 mm tube
Mass
Approx. 0.5 kg
Product name
Automatic calibration unit
*
Model
RA-ZC100V
Operating range
0.1 L/min to 0.5 L/min
Material of liquid contact section
PVC, EPDM, PVDF
Operating temperature
0
°
C to 50
°
C (no freezing)
Primary pressure
0.1 MPa to 0.2 MPa
Piping connection
sample inlet and outlet
φ
8 mm
×
φ
6 mm tube
Power supply
100 V AC, 50 Hz/60 Hz
Allowable power fluctuation range
90
%
to 110
%
of the power supply voltage
Power consumption
10 W
Mass
Approx. 1.5 kg
*: Use the automatic calibration unit indoors.
Product name
Manual calibration unit
Model
RA-ZCH
Operating range
0.1 L/min to 0.5 L/min
Material of liquid contact section
PVC, EPDM, PVDF
Operating temperature
0
°
C to 50
°
C (no freezing)
Primary pressure
0.1 MPa to 0.2 MPa
Piping connection
sample inlet and outlet
φ
8 mm
×
φ
6 mm tube
Mass
Approx. 1.0 kg