Product Handling Information
Operational precautions
Use of the equipment in a manner not specified by the manufacturer may impair the protection
provided by the equipment. And it may also reduce equipment performance.
z
Do not press any operation key with the tip of your nail or anything pointed.
z
Do not use solutions such as organic solvents.
z
When cleaning the measuring cell, be extremely careful not to scratch the brownish-red
area of the anode and reference electrode surfaces.
z
Direct sunlight will reduce the silver chloride covering the reference and anode electrode.
Install the sensor in a room to avoid direct sunlight.
z
Be sure not to scratch or contaminate the gold electrode section on the cathode surface.
If the surface of the gold electrode is scratched, it affects the reading.
z
When reattaching the bead cell, confirm that there are no beads inside the measuring cell
and around the O-ring, and then push it into the measuring cell.
z
Gradually open the supply source valve. If you open it abruptly, the flow rate or pressure
exceeds the specified level. In such a case, the beads may pop up from the bead cell.
Open the sampling valve slowly. Fully open valve passes excessive flow rate that kicks
out the ceramic beads from the bead cell. The adequate flow rate is 0.1 L/min to 0.5 L/
min.
z
Do not touch the rotating electrode.
Disposal of the product
When disposing of the product, follow the related laws and/or regulations of your country for
disposal of the product.
This product includes the lead battery of uninterruptible power supply system and the
tungsten lamp of the microscope light.