Troubleshooting
61
J
Troubleshooting for error codes
If an error code is displayed, take an appropriate action according to the following
troubleshooting list.
Error code
Occurrence condition
Cancellation
condition
Possible cause
Remedial action
E-11
(Zero
calibration
error)
Current exceeding
±
0.8
μ
A flows between A and
K during zero calibration
z
Cancelled
by
pressing
ESC
z
Cancelled
by
performing
recalibrati
on
Calibration has not been
performed correctly or
the detector itself is faulty
Once the adjustment value has
been stabilized, press ENT.
E-12
(Span
calibration
error)
Sensor sensitivity
deviates by the specified
ratio from the default
value during span
calibration
Calibration has not been
performed correctly or
the converter itself is
faulty
E-13
(Automatic zero
stability error)
Current exceeding
±
0.8
μ
A flows between A and
K during zero calibration
E-21
(Disconnection
of temperature
sensor)
When the resistance
value of the temperature
sensor exceeds 1.58 k
Ω
(when the temperature
sensor for 1 k
Ω
exceeds
150
°
C)
When the
conditions
described on
the left are
cancelled
Resistance error between
T and T of the
temperature sensor
If the resistance value is outside
the ranges described on the left,
the temperature sensor is faulty.
Replace the sensor with a new
one.
Wiring error of the
connector cable and
relay cable
Check that the wiring between T
and T is not open.
E-22
(Short circuit of
temperature
sensor)
When the resistance
value of the temperature
sensor is below 0.8 k
Ω
(when the temperature
sensor for 1 k
Ω
is below
−
50
°
C)
Resistance error between
T and T of the
temperature sensor
If the resistance value is outside
the ranges described on the left,
the temperature sensor is faulty.
Replace the temperature sensor
with a new one.
Wiring error of the
connector cable and
relay cable
Check that the wiring between T
and T is not short-circuited.
E-25
(Temperature is
outside the
measurement
range)
The temperature is
below
−
5
°
C or above
55
°
C
The wiring between T
and T is open (The lead
wire is disconnected)
Check the wiring.
Temperature sensor is
faulty
Check that the resistance
between T and T is within the
range from 1.00 k
Ω
to 1.21 k
Ω
.
The liquid temperature is
below
−
5
°
C or above
55
°
C
Adjust the liquid temperature
within the range from
−
5
°
C to
55
°
C.
E-27
(Temperature
calibration
error)
When
temperature
cal mode is
cancelled by
pressing
ESC
Temperature sensor is
faulty
Check that the resistance
between T and T is within the
range from 9.61 k
Ω
to 1.43 k
Ω
.