Operation
22
Operation
Preparation for operation
When you operate the instrument at the first time, be sure to complete the following
preparation.
z
Check of converter and detector installation condition. (Refer to “ Installation method ”
(page 11))
z
Check if the wirings are correct. (Refer to the “ Wire connection ” (page 16))
z
Are the terminal block screws firmly tightened?
z
Is the fluctuation of the power supply voltage within
±
10
%
of the rated voltage?
z
Check if the piping for mistake. (Refer to the “ Piping ” (page 12))
z
The piping and wiring for the solenoid valve to control zero calibration should be properly
carried out.
(Refer to “ Auto zero calibration ” (page 54))
z
The ceramic beads should be properly filled in the bead tub. (Refer to “ Cleaning/
replacing the bead filter A, bead filter and ceramic beads ” (page 50))
z
The sampling flow rate can be adjusted within 0.1 and 0.5 L/min. (Refer to “ Checking the
sample flow rate ” (page 47))
z
Initial setup (selection of transmission signal setting, operation condition settings for relay
contacts, etc.)
z
Calibration (zero and span) (Refer to “ Calibration ” (page 53))
J
Initial setup
The converter is shipped with initial settings.
If necessary, change the initial settings according to “ Setup menu ” (page 30).
For the initial setting values, refer to the table below.
Display
Item
Description
Setting
range
Initial
value
Refer
to page
PArA
Parameter setting
31
dF
Moving average count
(damping factor) setting
Increase dF to reduce the variations of the
measured data: Decrease dF to speed up the
response.
1 to 120
(times)
10
31
A.rEt
Auto return setting
This setting returns the system from the hold
mode to the meas mode automatically.
yES, no
yES
31
rEt.t
Auto return time setting
The hold mode will be back to the meas mode
automatically when the set time is passed from the
last key touch.
Note:
This is displayed when “Auto return setting” is set
to “yES”.
1 to 24 (h)
2
31
rEt.S
Transmission hold
cancellation time setting
This setting allows you to maintain the
transmission hold state until the measured value is
stabilized after the hold mode has been switched
to the meas mode. Set the time (in seconds) until
the maintained transmission hold state is
cancelled.
2 to 30 (s)
5
31
PH.Co
pH correction function
setting
Specify whether or not to apply the pH correction
to the measured residual chlorine value and how
to input the pH value.
inPut, oPt,
non
non
31