Measurement
33
Display setting of E-13
automatic zero
calibration attempt
error
Specify whether or not to display an automatic zero calibration
attempt error (E-13).
z
yES: Displays E-13 and also outputs a contact if zero calibration
fails because the value is not stabilized within the automatic zero
calibration time.
Note, however, that the transmission output for the next
measurement is not held even when E-13 is displayed and the
contact is output.
If the system judges that it is an error, measurement continues
under the condition of the latest zero calibration that has been
performed successfully.
E-13 can only be cancelled next time the zero calibration is
performed successfully or zero calibration is completed manually.
An error is more likely to occur when a longer zero stability time
is set. If the time is too short, zero calibration is performed with
chlorine concentration higher than zero. In this case, the
measured value may be lower than the actual value.
z
no: No error processing is performed and measurement
continues under the condition of the previous zero calibration.
yES, no
Contact output setting: R1 to R3
Tip
The other relays can be set in the same manner.
Reference
Refer to “ Control targets and their available options ” (page 37) for the options available for the
selected control target.
Contact output type
setting
Specify the contact output type.
z
CAL: Issues an output during calibration (including the hold time
after calibration)
z
HoLd: Issues an output in a transmission output hold state
z
t: Outputs a temperature alarm
z
AL: Outputs an alarm
z
CtrL: Issues a control output
z
non: Stops the function and no contact is output
CAL,
HoLd, t,
AL,
CtrL,
non
Control direction
setting
Specify whether to perform control using the upper limit operation or
lower limit operation.
z
Hi: Relay 1 is activated when a value is over the alarm level.
z
Lo: Relay 1 is activated when a value is under the alarm level.
Note:
This is displayed when “Contact output type setting” is set to “CtrL”,
“AL”, or “t”.
Hi, Lo
Control value setting
Specify the control value.
Note:
This is displayed when “Contact output type setting” is set to “CtrL”,
“AL”, or “t”.
0.00 to
3.00
(mg/L),
0.0 to
100.0
(
°
C)
Control width setting
Specify the control width.
Note:
This is displayed when “Contact output type setting” is set to “CtrL”.
0.02 to
1.00
(mg/L)
Display
Item
Description
Setting
range