Operation
26
E.in1
External input 1 setting
36
tyPE
Contact input type setting Specify the contact input type.
FLo, HoLd,
rng, CAL
FLo
36
FLo.in
Flow switch input function
setting
Specify whether to enable or disable the flow
switch input.
Note:
This is displayed when “Contact input type setting”
is set to “FLo”.
yES, no
no
36
FLo.t
Flow switch oFF delay
time setting
Specify the time until a flow rate error (E-61) is
displayed after the flow switch is turned OFF.
Note:
This is displayed when “Contact input type setting”
is set to “FLo”.
0 to 600 (s)
30
36
FLo.rt
Motor stop setting
Specify whether or not to stop the motor rotation
when a flow rate error (E-61) occurs.
Note:
This is displayed when “Contact input type setting”
is set to “FLo”.
yES, no
yES
36
E.in2
External input 2 setting
Note
When the settings of contact input 1 and 2 are the same, the input for contact input 2 is
disabled. However, the “EXT2” indicator on the screen lights up.
36
tyPE
Contact input type setting Contact input type setting
HoLd, rng,
CAL
rng
36
RS-485
address
RS-485 address setting
36
rS
RS-485 address setting
Specify the RS-485 communication address.
0 to 99
0
36
Display
Item
Description
Setting
range
Initial
value
Refer
to page