COMMISSIONING AND DECOMMISSIONING
Information for general use
Rev. M / 73023639
80
GasLab Q2
6.1.1.
Set password for Administrator (admin1)
Assign password at first login to the enCore device! In the delivery state, user passwords for
standard users are empty for all enCore devices, including admin1. Even if you create a new user
in enSuite, enSuite creates these users without a password.
We strongly recommend that each user changes his or her password
at the first log into the device.
Further details in sections
➪
System display Users (Login/Logout /
and
6.1.2.
Adapting the factory parameterization
In the factory parameterization (default parameter set), usually only calibration gas properties,
automatic calibration intervals and the purging / flushing times for the gases must be adjusted. For
extreme ambient conditions, the setting for the auxiliary housing heater or the fan is still required.
The default parameter set automatically switches on the additional heating system at 0°C
≙
32°F
≙
273K with a switching hysteresis of 2°C
≙
3,6°F
≙
2K. This supplementary heater function is only
available up to a maximum housing temperature of + 13°C
≙
55°F
≙
286K.
Other parameters are normally left at their factory-set values and, if necessary, supplemented by
customer-specific requirements.
To understand these parameters and to change their values if necessary, please refer to Chapter
8 Configuration and analysis software enSuite
enSuite online help
.
6.1.3.
Opening and closing the hardware parameter guard (SSW)
You can protect settings which affect the measuring properties. This is done with the Security Switch
(short: SSW) also called calibration switch. On delivery the connection is open. To open or close the
SSW, the device must be opened.
Never open installed devices without the permission and consent of the system operator.
Observe the explosion protection and safety regulations!
N'ouvrez jamais les appareils installés sans l'autorisation et le consentement de
l'opérateur du système. Respectez les règles de protection contre les explosions et de
sécurité !
Additionally, the housing hood may only be removed in controlled conditions
as described in ISO/IEC 61010-1.