Ethernet
Cable type refer to
5.4.2 Cables (power supply / communication)
≙
328ft.
The connection is made using the terminals designated TB3-1 to TB3-4 in the Q2 connection box.
The core colors depend on the standard used, normally EIA/TIA 568B. The twisted pairs must not be
untwisted or regrouped. The polarity of the cores within the core pair is unimportant and the core
pairs can even be swapped.
Core
pair
Plug
(pins)
Color depending on standard
Remark
Terminal
inside
Q2
EIA/TIA 568 B
EIA/TIA
568 A
IEC
REA
DIN 47.100
1
4 / 5
blue + white/blue
blue +
white
white/+
blue
white+
blue
white+
brown
2
3 / 6
white/green
+ green
white +
orange
red+
orange
turquoise
violet
green+
yellow
Only core
pairs 2 and 3
are required
TB3-3
TB3-4
3
1 / 2
white/orange
+ orange
white +
green
black+
grey
white+
orange
grey +
pink
TB3-1
TB3-2
4
7 / 8
white/brown +
brown
white +
brown
brown
tu
violet
blue +
red
The following drawing is only designed to illustrate the screening using an example.
Figure 5.7: Sample LAN/Ethernet connection (diagram)
Additional to the wiring, settings must be made before this interface can be used.
On the Q2
System display I/O (basic network settings / Inputs /Outputs)
Or with enSuite
Changing the IP address basic network and I/O settings