CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
Information for general use
Rev. M / 73023639
198
GasLab Q2
Requirements:
•
The software versions are independent from each other. Q2SENS-FW:1.2.0 is
essential for the update described here.
(See
7.3.5 Info display (Software Status)
•
The current software version must not be too old for the update. The basic
system requires version 3-08-A or newer.
(Perform the check described in section
•
The number of base calibration gases must also be correct.
(
Check by reading the existing parameter set and comparing the entries
with a new parameter set
).
You can use the change history of a new software version to check whether the new
functions, improvements, bug fixes and resolved security problems are relevant for
your enCore device.
The change history is available for software releases on the Elster-Instromet
website since 2020 as a PDF file in the download area of the respective device type.
.
www.elster-instromet.com/en/software-downloads
enSuite ensures the compatibility of the software modules. The exchangeable software modules
include the basic system (with its SFBs), all available AFBs as well as the optional approval file that
defines the access rights, e.g. for devices in fiscal use.
For all following steps you always need all current software versions
. The Elster-Instromet
website
www.elster-instromet.com/en/software-downloads
contains the latest software releases
for the Elster gas appliance series. Download the required version from the download area onto
the service PC and unpack it.
In most cases, a
new enSuite version
is required for a device release. Please check the enSuite
modification history on this website to find out which enSuite version is required or compatible. If
necessary, download this version and reinstall it or update your existing installation. The current
enSuite is available in the download area of the Elster-Instromet website:
The exchange of software modules requires the user login and should also be prevented by the
security switch (SSW), i.e. exchange is only possible with the switch open. Access rights should
only exist for the administrator or a user profile with the same rights. When opening the seal
switch, also observe the legal regulations of the respective country.