GASLAB Q2 AFTER DELIVERY AND AT THE PLACE OF USE
Information for general use
Rev. M / 73023639
78
GasLab Q2
Analog outputs
There are
four common electrically isolated active output circuits (“common ground”/short
-circuit-
resistant) with 0 or 4 to 20 mA in the connection box (TB7-terminals, details see figure 5.6). The
maximum supply voltage is around 9 V. The maximum load is 390
Ω
.
Cable type refer to
5.4.2 Cables (power supply / communication)
500m
≙
1640ft. The figure also shows the appropriate terminals in the connection box. The wiring
in the drawing is only designed to illustrate the screening using an example.
Figure 5.11: Sample connection of digital outputs
In addition to the wiring, various enSuite settings are required before this interface
can be used.
Changing the IP address basic network and I/O settings