GASLAB Q2 DISPLAYS (OPERATE THE DEVICE)
Information for general use
Rev. M / 73023639
106
GasLab Q2
Figure 7.21: System info
–
Software info display
The above figure only shows examples.
If, for example, the device is not used without an approval file, the “Secured
parameters” display will not be available. Furthermore, some links can only be
selected if there are no red entries. As the example shows, the component displays
do not have a uniform structure.
The “
Software status
” display shows the short name of an
approval file
in second place on the list.
This file contains the definition of which parameters are backed up before being changed so that
the device can be operated in compliance with customer specifications or approvals.
Operation without an approval file is also possible. In this case, as in the example above,
NonFiscal
is displayed as the abbreviation. If the device is operated with an
approval file
and a closed SSW,
the appropriate parameters are protected to prevent them being changed.
If you wish or have to backup parameters with an approval file, refer to section
6.1.3 Opening and closing the hardware parameter guard (SSW)
and section
8.4.10 Fiscal parameters and optional using approval file (Select approval)
If necessary, please contact Honeywell.
If an approval file is used and both the official access rights of the
parameterization as well as the fiscal software component versions are in line
with the approval file, then this text is black; the text is red in the event of any
deviations
.
If you activate the hyperlink of the
Approval,
a display is then evoked which contains
detailed information on the approval.
If the device is operated with an
approval file
and a closed SSW, the appropriate parameters are
protected to prevent them being changed.