EK280
Application Manual
45
6.3 Status output (signal output)
Each of the outputs 1 to 4 (terminals DA1 to DA4) can be programmed as a status output (signal output).
6.3.1 Parameterizing by keyboard
i
The use of
enSuite
parameterizing software is strongly recommended for parameterizing a status
output (see below).
Parameterization with the keyboard on the EK280 is described below.
a) Setting up the output mode:
−
Move the cursor to the tab "
Serv.
" and subsequently via the following path to the mode for the
corresponding output
"
Md.O1
"
to
"
Md.O4
"
:
Serv.
Outputs
Output 1
Md.O1
Serv.
Outputs
Output 2
Md.O2
Serv.
Outputs
Output 3
Md.O3
Serv.
Outputs
Output 4
Md.O4
i
The status outputs of the EK280 can be operated with active or inactive logic. The output mode
must be programmed accordingly for this purpose (see below).
−
Press the ENTER key.
⇒
The actual parameterized value flashes.
−
Press one of the arrow keys
or
until the
""
(Logic active: message is active, when the output is switched on.) or
"status-"
(Logic
inactive: message is active, when the output is switched off) flashes.
−
Press the ENTER key to confirm the value setting. Input can be canceled using the ESC key.
b) Setting up the status display
for the output:
−
When in the corresponding menu press the
key repeatedly until
"SpO1
",
"SpO2
",
"SpO3
", or
"SpO4
" is displayed.
−
Press the ENTER key.
⇒
The actual parameterized value flashes.
−
Move the cursor to the digits using the
or
arrow keys and update them using the
or
arrow
keys to a value specified in the table.