EK280
Application Manual
98
If the device is to connect with servers which are addressed via their network names, then also enter the
addresses for the domain name server (“
DNS12
”, “
DNS22
”).
These access details can be obtained from IT officer responsible for the system.
You can define time windows within which you can access the EK280 in the basic configuration which is now
set up. You can deactivate time windows which are not needed by setting the start and end of the time
window to the same value.
−
In the top part of the device window, open the path “
Parameters
” > “
Interfaces
” by clicking on the
relevant plus signs, and then select “
Call time windows
”.
−
In call window 3, set the start to “00:00” and the end to “23:59” and then select the “Transfer
changes to device” icon at the top of the enSuite window.
−
Switch on the mains power supply of the EK280 again. The plug symbol in the Main Menu of the
EK280 shows the external supply.
−
Wait for around a minute for the EK280 to detect the iCE280.
i
When the Ethernet module is active two LEDs flash on the module (
ACT
and
LINK
).
If the LEDs do not flash, either the module is not active (possible cause: the time window is still open) or
there is a fault in the wiring.
7.2.5 Retrieving data with enSuite
a) Opening the test time window for the following test retrievals with enSuite
i
The test time window only needs to be open if time windows 3 and 4 are closed.
−
If there is no communication link to the EK280 via the optical interface, it must be established.
(
3.1 Setting up the communication link to the EK280 via optical interface
−
In the top part of the device window, open the path “
Parameters
” > “
Interfaces
” by clicking on the
relevant plus signs, and then select “
internal modem
” (not the plus sign in front).
−
Click on the value “
0
” next to “
Test call acceptance window
” in the right-hand part of the window and
enter a duration in minutes (e.g. 10) for which the test call acceptance window should remain open
for the test retrieval. For a battery-operated EK280, do not enter an excessively long duration, as this
will adversely affect the battery life.