Document No: C-61-00003-3
Figure 8-2 Optical Window Adjuster
d) Click connect button after UAM is detected and COM port number is displayed. Click search, if COM
port number is not displayed.
e) Click Start Adjustment button when it is enabled.
When application is ready, password window will appear. Enter the password of the device and click
“OK” button. UAM will restart once and start the calibration. It will take approximately 6 minutes to
complete the calibration of 8 contamination channels. 7-segment display on UAM will show the code “CE”
during the process.
Figure 8-3 Adjustment completed display Figure 8-4 Error display
g) When adjustment is complete, dialog box as shown in Figure 8-3 will appear. Click the OK button and
then click disconnect button (Figure 8-2). Remove the USB cable. UAM will switch to normal operation.
Retry the adjustment after checking the optical window and USB connection if an error dialog as shown in
Figure 8-4 is displayed. If adjustment fails even after repeated trial contact the nearest distributor or sales
representative for support. Send the log file and calibration data file for investigating the cause of failure
in such case.
Содержание UAM-05LP
Страница 1: ... Document No C 61 00003 3 ...
Страница 104: ... Document No C 61 00003 3 Figure 7 33 b Function Figure7 33 c Area ...
Страница 107: ... Document No C 61 00003 3 Figure 7 36 a Project report tab Figure 7 36 b Error report tab ...
Страница 148: ... Document No C 61 00003 3 13 External dimension 13 1 UAM 05LP ...
Страница 149: ... Document No C 61 00003 3 13 2 Base mounting bracket ...
Страница 150: ... Document No C 61 00003 3 13 3 Rear mounting bracket ...
Страница 151: ... Document No C 61 00003 3 13 4 Cover Protection Bracket ...
Страница 152: ... Document No C 61 00003 3 14 EC Declaration of conformity ...
Страница 153: ... Document No C 61 00003 3 ...