Alarms and error messages
83
SAVING ALARMS
All alarms are saved in the device, including date, alarm time, alarm cause,
and the measured values at the time of the alarm. The memory depth is
approximately 1 year.
READING OUT THE ALARM MEMORY ON THE DEVICE
You can view the last 200 alarm messages in the alarm screen. To do this,
proceed as follows:
1. Activate the alarm screen using the cursors or .
2. Select “Alarm” using the cursor or the touch wheel.
3. Press the OK button. The saving location of the alarms and the total
number of alarms are highlighted with a black bar.
4. Select the desired alarm using the cursors or , in order to view the
alarm time, alarm cause and measured values at the time of alarm.
FORWARDING ALARMS
Alarms can be forwarded by means of a nurse call or the optionally available
remote alarm box. This allows even better monitoring of the device to be
achieved in the home or clinic. The use of the remote alarm box or a nurse
call is especially recommended when several ventilators are used in one
room, as this allows the device generating the alarm to be easily identi
fi
ed.
Fig. 45: Reading out alarm memory
Fig. 46: Remote alarm box
ALARM