68
Operating the device
SETTING THE SOFTSTART TIME
The use of the “Softstart” function can bring some relief if the patient has
not really become accustomed to ventilation.
The softstart function slowly increases the pressure in the time programmed
by you starting with a de
fi
ned initial ramp pressure, up to the prescribed
pressure.
Ventilation mode
Starting ramp pressure
IPAP
EPAP
CPAP
Valve tube system: 0 hPa
Leakage tube system: 2 hPa
---
(A)PCV, PSV(-S)
4 hPa
0 hPa
S, T, ST
4 hPa
2 hPa
Table 9: Starting ramp pressures
1. Activate the comfort screen using the cursors or .
2. Select “Softstart” using the cursor or the touch wheel.
3. Press the OK button.
4. Set the desired time using the cursors or or the touch wheel. The
maximum time is preset by the physician.
5. Con
fi
rm the settings using the OK button.
IPAP
0
IPAP
IPAP
4
PEEP
PEEP
4
2
2
0
0
Softstart
Time [min]
Time [min]
Time [min]
Softstart
Softstart
Pressure [hPa]
CPAP mode
(A)PCV/PSV(S) - mode
S-/T-/ST - mode
Pressure [hPa]
Pressure [hPa]
Valve tube system
Leakage
tube system
Fig. 34: Setting the softstart time
Содержание TRENDvent
Страница 103: ...Notes 103 NOTES ...