Commissioning
53
USING OXYGEN
Before using oxygen, the safety instructions as of page 16 must be read.
The supply of oxygen is possible in all ventilation modes. Please note that
changes to the ventilation parameters, such as e.g. respiratory pressure,
I:E, respiratory frequency, will lead to a change in FiO
2
content.
Oxygen may only be supplied during active ventilation.
SUPPLYING OXYGEN
1. Connect the tube from the oxygen source to the oxygen connection
adapter.
2. Insert the oxygen connection adapter into the oxygen connection ( ).
Only the oxygen connection adapter supplied may be used to
connect oxygen. Otherwise, there is a risk that the back-stop in
the connection is damaged.
3. Switch the device on.
4. Start ventilation and wait for several respiratory cycles.
5. Start supplying the oxygen.
STOPPING THE SUPPLY OF OXYGEN
1. Stop the supply of oxygen at the oxygen source.
2. Allow ventilation to continue for a number of respiratory cycles.
3. Stop ventilation.
Fig. 22: Oxygen connection
1
2