Setting Threshold Values for the Comparator
7
Set the voltage evaluation
method.
When you have chosen the
[PoL]
, reversely connecting the
red and black wires of the test
lead will cause the
[WARNING]
result.
(Black)
(Red)
(Red)
(Black)
[WARNING]
Voltage
evaluation
method
Comparison method
Data to be
saved
[AbS]
(Default
setting)
Evaluates absolute voltage
values, regardless of whether
they are positive or negative.
Signed
(minus sign
only)
[PoL]
The
[WARNING]
result will be
given for the negative voltage
value.
If the test lead is placed into
contact in reverse polarity with
a battery (the red and black
probes with the negative and
positive terminals, respectively),
the
[WARNING]
result will be
given.
Signed
(minus sign
only)
You can set the evaluation method using version 1.8 or
later of GENNECT Cross.
HIOKI BT3554F961-00
HIOKI BT3554F961-00
Содержание BT3554-50
Страница 2: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 14: ...vi Contents HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 94: ...80 Canceling the Comparator Function HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 126: ...112 Excel Direct Entry Function HID Connection HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 183: ...169 Index Z Zero adjustment 50 Zero adjustment board 51 HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 184: ...170 Index HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 185: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 186: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 187: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...