132
Function Specifications
(13) Automatic power saving
Operation
Automatically turns off the instrument if inactivity
continues or a measurement-current anomaly continues
to be detected for 10 minutes (±1 minute).
Display
The
[APS]
segment appears.
Default setting
Enabled
Conditions
where the
function is
disabled
(The function is
automatically
disabled)
During communications
While the measurement recording prompt is operating
How to set
See “1.4 Names and Functions of Parts” (p. 28).
(14) Instrument’s battery level detection
Operation
Displays instrument’s battery level
(The accuracy can be guaranteed until the
indicator blinks)
Display
Four-level indicator (For LR6 Alkaline batteries).
10.1 V or more
9.2 V to 10.1 V
8.0 V to 9.2 V
(blinking) 7.6 V to 8.0 V
(Shutdown)
Less than 7.6 V
The indicator has an error of ±0.2 V.
How to disable
None
(15) Date and time
Operation
24-hour clock; leap years are adjusted automatically.
Accuracy
Approx. 4 minutes per month
HIOKI BT3554F961-00
HIOKI BT3554F961-00
Содержание BT3554-50
Страница 2: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 14: ...vi Contents HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 94: ...80 Canceling the Comparator Function HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 126: ...112 Excel Direct Entry Function HID Connection HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 183: ...169 Index Z Zero adjustment 50 Zero adjustment board 51 HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 184: ...170 Index HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 185: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 186: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 187: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...