149
Replacing the Tip Pin of the Test Lead
4
Pinch the transparent tube and pull the pin base to remove
the tip pin.
Transparent tube
Pin base
Do not subject the cable to stress to prevent a wire brake.
5
Insert a new 9465-90 Tip Pin and press the tip against a
hard surface.
6
Assemble the grip in the reverse order of disassembling.
• To prevent a wire break, project the bush about
1 mm from the end face of the collet.
• Be careful not to push the bush too deeply.
Collet
Bush
Do not pull or twist the cable during assembling.
7
After tightening the collet nut, gently pull and twist the
cable to check if it has firmly been fixed. Loose connections
could cause a wire break or poor contact.
8
Check for proper operation.
Measure a non-defective object to check if the measured
resistance is correct before use.
HIOKI BT3554F961-00