126
Function Specifications
(4) Noise-frequency reduction
(Noise reduction technology)
Operation
Avoids noise frequencies using the noise reduction
technology.
Display
Function enabled:
The
segment appears.
During reduction of noise frequencies:
The
segment blinks.
Default setting
Disabled
How to set
See “3.3 Noise-Frequency Reduction Function”
(5) Zero adjustment
Operation
Internally adjusts the instrument so that it provides the
zero indications corresponding to the measured values
obtained through zero adjustment.
Display
During zero adjustment:
The
[0ADJ]
segment blinks (While obtaining the
correction value).
Enabled:
The
[0ADJ]
segment appears (With the correction
value enabled).
Default setting
Disabled
Correction range
Up to 300 counts for each range of the resistance and
voltage values
How to correct
See “3.4 Adjusting the Zero Point (Zero Adjustment)”
HIOKI BT3554F961-00
HIOKI BT3554F961-00
Содержание BT3554-50
Страница 2: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 14: ...vi Contents HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 94: ...80 Canceling the Comparator Function HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 126: ...112 Excel Direct Entry Function HID Connection HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 183: ...169 Index Z Zero adjustment 50 Zero adjustment board 51 HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 184: ...170 Index HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 185: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 186: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 187: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...