Effects of Current Density
To avoid these effects, detect voltage across the two points located
the inside of the two current injecting points using a four-terminal
test lead such as the Hioki 9453 Four Terminal Lead. The current
distribution is generally considered to become stable at a distance
equivalent to the width or thickness or more inwardly away from the
current injecting points.
Inject
current
Inject
current
Figure 1: Equipotential lines on a metal plate, showing the current
distribution at 50 µV intervals when a current of 1 A is injected
from the distal points of the plate (W300 × L370 × t0.4)
As shown in Figure 2, it is preferable to place the SENSE terminals
at the points a distance equivalent to the width or thickness inwardly
away from the SOURCE terminals are placed.
SOURCE
SOURCE
SENSE
SENSE
Greater or equal to W or t Greater or equal to W or t
W
t
Figure 2: Contact locations when the object under measurement is wide
or thick
HIOKI BT3554F961-00
HIOKI BT3554F961-00
Содержание BT3554-50
Страница 2: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 14: ...vi Contents HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 94: ...80 Canceling the Comparator Function HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 126: ...112 Excel Direct Entry Function HID Connection HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 183: ...169 Index Z Zero adjustment 50 Zero adjustment board 51 HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 184: ...170 Index HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 185: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 186: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Страница 187: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...