65
────────────────────────────────────────────────────
7.4 Commands Specific to the 3157
────────────────────────────────────────────────────
:UNIT
■
Sets the unit of the maximum and minimum test values.
:UNIT?
■
Queries the unit of the maximum and minimum test values.
Syntax
:UNIT
_<data>
<data>
OHM: resistance (character data)
VOLT: voltage (character data)
Function
Sets the unit of the maximum and minimum test values.
Note
If the unit of the maximum and minimum test values is set to "OHM"
(resistance), the maximum and minimum test values (resistance) settings are
reflected in the test.
If the unit of the maximum and minimum test values is set to "VOLT"
(voltage), the maximum and minimum test values (voltage) settings are
reflected in the test.
Example
Transmission
:UNIT OHM
The unit of the maximum and minimum test values is set to OHM.
Error
If <data> is other than character data and numerical value described above, a
command error occurs.
The execution of this command in a state other than the following states
causes an execution error:
・
READY state
・
Optional function setting screen
Syntax
:UNIT?
Function
Returns the setting for the unit of the maximum and minimum test values as
character data.
<data>
OHM/VOLT (character data)
Example
Response
If headers are on
:UNIT OHM
If headers are off
OHM
Error
If the response message is longer than 300 bytes, a query error is generated.