34
────────────────────────────────────────────────────
7.3 Particular Commands
────────────────────────────────────────────────────
Voltage or resistance indicator (unit)
Resistance
Test time enablement
ON
Maximum test value enablement
ON
Minimum test value enablement
OFF
Output current value
25.0 A
Maximum test value (resistance)
0.100
Ω
Minimum test value (resistance)
0.000
Ω
Maximum test value (voltage)
2.50 V
Minimum test value (voltage)
0.00 V
Test time
60.0 s
*
RST
■
Performs device initial setting.
*
TST?
■
Requests execution of, and queries the result of, the self test.
Syntax
*
RST
Function
Resets the 3157. The items which are reset are listed below.
Error
If the data parameters are set after this command, a command error occurs.
Syntax
*
TST?
Function
・
Performs the self test of the 3157, and returns the result thereof as a
numerical value in NR1 format between 0 and 3.
・
No header is affixed to the response message.
Bit 0: A ROM error occurred.
Bit 1: A RAM error occurred.
Bits 2 to 7: Unused
Example
Response
3
A ROM error (bit 0) and a RAM error (bit 1) have occurred.
Error
If the response message is longer than 300 bytes, a query error occurs.
The execution of this command in a state other than the READY state causes
an execution error.