41
────────────────────────────────────────────────────
7.4 Commands Specific to the 3157
────────────────────────────────────────────────────
:CONFigure:VLOWer
■
Sets the minimum test value (voltage).
:CONFigure:VLOWer?
■
Queries the minimum test value (voltage).
Syntax
:CONFigure:VLOWer
_<data>
<data>
Numerical data in NR2 format between 0.00 and 6.00
Function
Sets the minimum test value (voltage).
The numerical value can be in NRf format, but rounding is performed for
figures beyond the last valid decimal place.
Note
The minimum test value (voltage) can be used as the basis for the test only
when the maximum and minimum test values are set to be expressed in
"VOLT" (voltage) and when the minimum test value setting is not disabled,
and when "ON" is selected in the ON/OFF setting for the minimum test value.
Example
Transmission
:CONFigure:VLOWer 0.00
The minimum test value (voltage) is set to 0.00 V.
Error
If <data> is other than numerical value described above, an execution error
occurs.
The execution of this command in a state other than the READY state causes
an execution error.
Syntax
:CONFigure:VLOWer?
Function
Returns the minimum test value (voltage) setting as a numerical value in NR2
format.
<data>
Numerical data in NR2 format between 0.00 and 6.00
Example
Response
If headers are on
:CONFIGURE:VLOWER 0.00
If headers are off
0.00
Error
If the response message is longer than 300 bytes, a query error is generated.